Fault Self‑Diagnosis Library

Troubleshoot common problems before calling for service. Most issues can be resolved with simple checks.

❌ Zero‑Point Drift

If your scale reads non‑zero when empty, check:

  • Mechanical binding – debris under the platform
  • Load cell damage – visual inspection for cracks
  • Moisture ingress – check junction box seals
Need help? Contact support →

⚠️ Unstable Readings

If weight fluctuates, check:

  • Loose cable connections at indicator or junction box
  • Electrical interference – keep cables away from motors
  • Corner adjustment – check individual cell outputs
Need help? Contact support →

📟 Indicator Error Codes

Common XK3190/DS3 indicator errors:

  • Err 01: Load cell signal out of range – check wiring
  • Err 02: ADC failure – call for service
  • Err 03: Zero calibration required – re‑calibrate
Need help? Contact support →

📡 Communication Failure

If weight data is not reaching your PLC/SCADA:

  • Check RS485/Modbus wiring – polarity and termination
  • Verify baud rate and parity settings
  • Test with a laptop serial monitor
Need help? Contact support →